Performance evaluation of novel fluorescent-based lateral flow immunoassay (LFIA) for rapid detection and quantification of total anti-SARS-CoV-2 S-RBD binding antibodies in infected individuals

  • Farah M. Shurrab
    Affiliations
    Biomedical Research Center, Qatar University, 2713, Doha, Qatar
    Search for articles by this author
  • Nadin Younes
    Affiliations
    Biomedical Research Center, Qatar University, 2713, Doha, Qatar

    Department of Biomedical Science, College of Health Sciences, Member of QU Health, Qatar University, 2713, Doha, Qatar
    Search for articles by this author
  • Duaa W. Al-Sadeq
    Affiliations
    Biomedical Research Center, Qatar University, 2713, Doha, Qatar

    College of Medicine, Member of QU Health, Qatar University, 2713, Doha, Qatar
    Search for articles by this author
  • Na Liu
    Affiliations
    Shenzhen Mindray Bio-Medical Electronics Co., Ltd., Shenzhen, China
    Search for articles by this author
  • Hamda Qotba
    Affiliations
    Department of Clinical Research, Primary Health Care Centers, 26555, Doha, Qatar
    Search for articles by this author
  • Laith J. Abu-Raddad
    Affiliations
    Infectious Disease Epidemiology Group, Weill Cornell Medicine-Qatar, Cornell University, Qatar Foundation—Education City, 24144, Doha, Qatar

    World Health Organization Collaborating Centre for Disease Epidemiology Analytics on HIV/AIDS, Sexually Transmitted Infections, and Viral Hepatitis, Weill Cornell Medicine-Qatar, Cornell University, Qatar Foundation-Education City, 24144, Doha, Qatar

    Department of Healthcare Policy and Research, Weill Cornell Medicine, Cornell University, New York, NY, 10065, USA
    Search for articles by this author
  • Gheyath K. Nasrallah
    Correspondence
    Correspond to: Gheyath K. Nasrallah, Department of Biomedical Science, College of Health Sciences, Qatar University, Doha, Qatar, Women's Science building, C01, Tel: +974 4403 4817, Fax: +974-4403-1351, P.O. Box: 2713.
    Affiliations
    Biomedical Research Center, Qatar University, 2713, Doha, Qatar

    Department of Biomedical Science, College of Health Sciences, Member of QU Health, Qatar University, 2713, Doha, Qatar
    Search for articles by this author
Open AccessPublished:February 26, 2022DOI:https://doi.org/10.1016/j.ijid.2022.02.052

      Highlights

      • Finecare fluorescence lateral flow immunoassay (LFIA) test quantitatively detects SARS-CoV-2 spike protein receptor-binding domain (S-RBD) total antibodies.
      • Finecare test showed 92% sensitivity and 100% specificity.
      • Finecare test strongly correlates with surrogate virus-neutralizing test (sVNT) and automated assays.
      • Finecare test can be used to predict the amount of neutralizing antibodies.

      Abstract

      Background

      A vast majority of the commercially available lateral flow immunoassay (LFIA) is used to detect SARS-CoV-2 antibodies qualitatively. Recently, a novel fluorescence-based lateral flow immunoassay (LFIA) test was developed for quantitative measurement of the total binding antibody units (BAUs) (BAU/mL) against SARS-CoV-2 spike protein receptor-binding domain (S-RBD).

      Aim

      This study aimed to evaluate the performance of the fluorescence LFIA FinecareTM 2019-nCoV S-RBD test along with its reader (Model No.: FS-113).

      Methods

      Plasma from 150 reverse trancriptase–PCR (RT-PCR)-confirmed positive individuals and 100 prepandemic samples were tested by FincareTM to access sensitivity and specificity. For qualitative and quantitative validation of the FinCareTM measurements, BAU/mL results of FinCareTM were compared with results of 2 reference assays: the surrogate virus-neutralizing test (sVNT, GenScript Biotech, USA) and the VIDAS®3 automated assay (BioMérieux, France).

      Results

      FinecareTM showed 92% sensitivity and 100% specificity compared with PCR. Cohen's Kappa statistic denoted moderate and excellent agreement with sVNT and VIDAS®3, with values being 0.557 (95% CI: 0.32–0.78) and 0.731 (95% CI: 0.51–0.95), respectively. A strong correlation was observed between FinecareTM/sVNT (r = 0.7, p < 0.0001) and FinecareTM/VIDAS®3 (r = 0.8, p < 0.0001).

      Conclusion

      FinecareTM is a reliable assay and can be used as a surrogate to assess binding and neutralizing antibody response after infection or vaccination, particularly in none or small laboratory settings.

      Keywords

      1. Introduction

      Soon after the emergence of SARS-CoV-2 in December 2019 and its declaration as a pandemic by the World Health Organization (WHO), the need for accurate, sensitive, and rapid detection of SARS-CoV-2 for the control and prevention of the disease was urgent (
      • Theel ES
      • Slev P
      • Wheeler S
      • Couturier MR
      • Wong SJ
      • KJJocm Kadkhoda
      The role of antibody testing for SARS-CoV-2: is there one?.
      ). Several commercial COVID-19 test kits were developed in response to this urgent need for the detection of either nucleic acid or antibodies (
      • Van Walle I
      • Leitmeyer K
      • Broberg EKJE.
      Meta-analysis of the clinical performance of commercial SARS-CoV-2 nucleic acid and antibody tests up to 22 August 2020.
      ). Although reverse trancriptase–PCR (RT-PCR) is used as a gold standard test, it requires specialized conditions, expensive equipment, and qualified personnel for sampling and testing. These limitations pose a challenge in a pandemic setting, which requires rapid and reliable tests that can be used to screen populations (
      • Dortet L
      • Ronat J-B
      • Vauloup-Fellous C
      • Langendorf C
      • Mendels D-A
      • Emeraud C
      • et al.
      Evaluating 10 commercially available SARS-CoV-2 rapid serological tests by use of the STARD (Standards for Reporting of Diagnostic Accuracy Studies) method.
      ). Thus, serological testing, particularly point‐of‐care approaches such as lateral flow immunoassay (LFIA), provides a rapid, portable, and cost effective method that is complementary to RT-PCR. In addition, serological tests have important applications in public health; they can be used in epidemiological survey studies, detect past infections, and determine herd immunity in the community. Furthermore, serological tests have a vital role in convalescent plasma treatment and determining vaccine effectiveness (
      • Nadoushan MJ
      • Ahmadi S
      • Nadoushan PJJIJoP
      Serology testing for SARS-CoV-2: Benefits and challenges.
      ).
      Conventional LFIA used in clinical settings for diagnosing SARS-CoV-2 infection are done manually within 15 minutes, and the results are interpreted qualitatively by the naked eye. As a result, errors due to manual operation and inadequate visual sensitivity interpretation can occur. Recently, Wondfo Biotech developed fluorescence-based lateral flow immunoassay (LFIA) to detect total anti-S-RBD binding antibodies (binding antibody unit [BAU]) against SARS-CoV-2. Finecare™ 2019-nCoV RBD Antibody Test is a fluorescence immunoassay that is carried out semiautomatically along with a small portable device. The combination of fluorescence and LFIA provides higher sensitivity, allows quantitative detection of antibodies, and is easily affordable and accessible in small clinical laboratories, research settings, or point-of-care testing after infection or vaccination.
      To the best of our knowledge, there is only 1 report in the literature evaluating the rapid fluorescence‐based LFIA immunoassays, which aimed to validate the performance of rapid SARS‐CoV‐2 IgM and IgG test kits based on fluorescence immunochromatography (
      • Kang K
      • Huang L
      • Ouyang C
      • Du J
      • Yang B
      • Chi Y
      • et al.
      Development, performance evaluation, and clinical application of a Rapid SARS-CoV-2 IgM and IgG Test Kit based on automated fluorescence immunoassay.
      ). In this study, we aimed to evaluate the performance of a rapid semiautomated fluorescence-based LFIA. This assay is designated to provide qualitative and quantitative measurements of SARS‐CoV‐2 anti-S-RBD total antibodies. One of the biggest advantages of this assay over the commercially available LFIAs is that not only are the results quantitative but they are also converted to BAU/mL (instead of AU/mL) as recently recommended by WHO.

      2. Materials and Methods

      2.1 Study design and ethical approval

      We evaluated the performance of FinecareTM 2019-nCOV Antibody test and its reader (Model No.: FS-113) (Guangzhou Wondfo Biotech Co., China) (from here on abbreviated as FinecareTM). The serum samples were collected from volunteer individuals between July 26 and September 9, 2020 as a nationwide survey substudy in Qatar (
      • Al-Thani MH
      • Farag E
      • Bertollini R
      • Al Romaihi HE
      • Abdeen S
      • Abdelkarim A
      • et al.
      Seroprevalence of SARS-CoV-2 infection in the craft and manual worker population of Qatar.
      ,
      • Syed MA
      • Al Nuaimi AS
      • Nasrallah GK
      • Althani AA
      • Yassine HM
      • Zainel AA
      • et al.
      Epidemiology of SARS-CoV2 in Qatar's primary care population aged 10 years and above.
      ). The clinical symptoms of the samples varied between asymptomatic and symptomatic cases (S1 table). The project was approved by the institutional review board of Qatar University (QU-IRB 1492-E/21 and QU-IRB 1469-E/21). These samples were used in previous studies (
      • Al-Jighefee HT
      • Yassine HM
      • Nasrallah GKJP
      Evaluation of antibody response in symptomatic and asymptomatic COVID-19 patients and diagnostic assessment of new IgM/IgG ELISA kits.
      ,
      • Yassine HM
      • Al-Jighefee H
      • Al-Sadeq DW
      • Dargham SR
      • Younes SN
      • Shurrab F
      • et al.
      Performance evaluation of five ELISA kits for detecting anti-SARS-COV-2 IgG antibodies.
      ,
      • Younes S
      • Al-Jighefee H
      • Shurrab F
      • Al-Sadeq DW
      • Younes N
      • Dargham SR
      • et al.
      Diagnostic efficiency of three fully automated serology assays and their correlation with a novel surrogate virus neutralization test in symptomatic and asymptomatic SARS-COV-2 individuals.
      ).

      2.2 Sensitivity and specificity determination

      The sensitivity of FinecareTM was determined using sera from 150 RT-PCR–confirmed individuals (7–>21 days). Qiagen extraction kit was used to extract RNA from nasopharyngeal swab specimens. Then, the SuperscriptIII One-Step RT-PCR kit was used to test for SARS-CoV-2 according to manufacturer's instruction (Cat No. 12594100, ThermoFisher, USA) (
      • Yassine HM
      • Al-Jighefee H
      • Al-Sadeq DW
      • Dargham SR
      • Younes SN
      • Shurrab F
      • et al.
      Performance evaluation of five ELISA kits for detecting anti-SARS-COV-2 IgG antibodies.
      ). Samples were tested using 2 sets of primers targeting the E gene and confirmed with 2 different sets of primers targeting the RdRp gene (
      • Corman VM
      • Landt O
      • Kaiser M
      • Molenkamp R
      • Meijer A
      • Chu DK
      • et al.
      Detection of 2019 novel coronavirus (2019-nCoV) by real-time RT-PCR.
      ,
      • Yassine HM
      • Al-Jighefee H
      • Al-Sadeq DW
      • Dargham SR
      • Younes SN
      • Shurrab F
      • et al.
      Performance evaluation of five ELISA kits for detecting anti-SARS-COV-2 IgG antibodies.
      ). Complete descriptions for these participants are summarized in Table S1. The specificity was examined using 100 prepandemic plasma samples collected before 2019, which were used in previous studies (
      • Nasrallah GK
      • Dargham SR
      • Mohammed LI
      • Abu-Raddad LJJJomv
      Estimating seroprevalence of herpes simplex virus type 1 among different Middle East and North African male populations residing in Qatar.
      ,
      • Nasrallah GK
      • Dargham SR
      • Sahara AS
      • Elsidiq MS
      Abu-Raddad LJJJoCV. Performance of four diagnostic assays for detecting herpes simplex virus type 2 antibodies in the Middle East and North Africa.
      ,
      • Smatti MK
      • Nasrallah GK
      • Al Thani AA
      • Yassine HMJV
      Measuring influenza hemagglutinin (HA) stem-specific antibody-dependent cellular cytotoxicity (ADCC) in human sera using novel stabilized stem nanoparticle probes.
      ). The panel included plasma samples seropositive for (a) dengue virus (n = 26), (b) parvovirus B19 (n = 8), and (c) nonrespiratory viruses (n = 66).

      2.3 Serological assays

      2.3.1 Finecare™ 2019-nCoV RBD Antibody Test

      Finecare™ 2019-nCoV RBD Antibody Test is a fluorescence immunoassay technology, specifically the sandwich immunodetection method. It uses fluorescently labeled SARS-CoV-2 S-RBD proteins to form immune complexes by binding to SARS-CoV-2 S-RBD antibodies present in the specimen. It is used along with the Finecare™ FIA Meters (Model No.: FS-113) for the quantitative detection of total antibodies against SARS-CoV-2 S-RBD (Co.). The test can be done using various specimens, including fingerstick whole blood, venipuncture blood, serum, or plasma. There are 2 test modes for Finecare™ FIA Meters: standard test mode and quick test mode. The difference is that samples are incubated inside the cartridge holder of Finecare™ FIA Meters in the standard mode, whereas the samples are incubated at room temperature in the quick mode. The test was done according to the manufacturer's instructions. In brief, 20 µL of plasma was added to the provided buffer tube and mixed for 45 seconds. Then, 75 µL was added to the test cartridge, incubated for 15 minutes at room temperature (RT), and then inserted in the test cartridge holder of Finecare™ FIA Meters holder for measurement on quick mode. Finecare™ FIA Meter displays the test result automatically on the screen. The results are given as relative fluorescence units (RFU, arbitrary units [AU]/mL). This test has a WHO international standardization factor to convert readings to from AU/mL to binding antibody units (BAU)/mL (1 AU/mL = 20 BAU/mL). Readings ≥1 AU/mL or ≥20 BAU/ml indicate positive results.

      2.3.2 cPass GeneScript sVNT

      Food and Drug Administration (FDA)-approved SARS-CoV-2 surrogate virus neutralization test (sVNT) developed by GeneScript is widly used in the literature as a reference assay to measure neutralizing antibodies (Cat. No. L00847-C, GenScript Biotech, USA) (
      • Ismail A
      • Shurrab FM
      • HT Al-Jighefee
      • Al-Sadeq DW
      • Qotba H
      • Al-Shaar IA
      • et al.
      Can commercial automated immunoassays be utilized to predict neutralizing antibodies after SARS-CoV-2 infection? A comparative study between three different assays. Front Biosci.
      ,
      • Meyer B
      • Reimerink J
      • Torriani G
      • Brouwer F
      • Godeke G-J
      • Yerly S
      • et al.
      Validation and clinical evaluation of a SARS-CoV-2 surrogate virus neutralisation test (sVNT).
      ,
      • Younes S
      • Al-Jighefee H
      • Shurrab F
      • Al-Sadeq DW
      • Younes N
      • Dargham SR
      • et al.
      Diagnostic efficiency of three fully automated serology assays and their correlation with a novel surrogate virus neutralization test in symptomatic and asymptomatic SARS-COV-2 individuals.
      ). This assay detects antibodies that inhibit the interaction between SARS-CoV-2 S-RBD–horseradish peroxidase (HRP) fusion protein and ACE2 that is coated in a 96-well plate (
      • Meyer B
      • Reimerink J
      • Torriani G
      • Brouwer F
      • Godeke G-J
      • Yerly S
      • et al.
      Validation and clinical evaluation of a SARS-CoV-2 surrogate virus neutralisation test (sVNT).
      ). The assay highly corelates with conventional pseudovirus neutralization tests (pVNTs, R2 = 0.84) and showed high specificity (99.9%) and sensitivity (100%) (
      • Tan CW
      • Chia WN
      • Qin X
      • Liu P
      • Chen MI-C
      • Tiu C
      • et al.
      A SARS-CoV-2 surrogate virus neutralization test based on antibody-mediated blockage of ACE2–spike protein–protein interaction.
      ). The test was done according to the manufacturer's instructions. An inhibition value of ≥30% signal inhibition was considered positive and <30% signal inhibition was considered negative. The WHO international standardization factor for this assay was used to convert readings from percent inhibition to IU/mL by applying the recently published formula (
      • Zhu F
      • Althaus T
      • Tan CW
      • Costantini A
      • Chia WN
      • Chau NVV
      • et al.
      WHO international standard for SARS-CoV-2 antibodies to determine markers of protection.
      ).

      2.3.3 BioMérieux VIDAS®3

      VIDAS®3 SARS-CoV-2 IgG (REF 424114, BioMérieux, France) is a widely used automated CE-marked in vitro diagnostic (IVD) assay, which was granted the emergency use authorization (EUA) by FDA in early 2021 The assay principle is based on a 2-step enzyme immunoassay combined with an enzyme-linked fluorescent assay (ELFA) detection technique. The test uses a solid-phase receptacle (SPR) coated with SARS-CoV-2 S-RBD of the spike protein. The test was done according to the manufacturer's instructions. In brief, the test was calibrated with standard (S1), a positive control (C1), and a negative control (C2). Then, 100 µL of the sample was added to the test strip. The results are generated as relative fluorescence value (RFV) and automatically calculated by the instrument; according to the S1 standard and sample RFV, an index value (i) is obtained (where i = RFVsample/RFVS1). The test is interpreted as negative when i < 1.00 and positive when i ≥ 1.00 (
      • Renard N
      • Daniel S
      • Cayet N
      • Pecquet M
      • Raymond F
      • Pons S
      • et al.
      Performance Characteristics of the Vidas SARS-CoV-2 IgM and IgG Serological Assays.
      ). All readings were standardized to BAU/mL by applying the WHO international standard (20.33 BAU/mL) for the VIDAS®3 SARS-CoV-2 IgG.

      2.4 Statistical analysis

      Using RT-PCR as the reference test, sensitivity, specificity, overall percent agreement, and Cohen's Kappa coefficient (κ) were calculated to assess the performance of FinecareTM. κ measures inter-rater reliability as well as the likelihood that an agreement will occur by chance (
      • Ben-David AJESwA
      Comparison of classification accuracy using Cohen's Weighted Kappa.
      ,
      • McHugh MLJBm
      Interrater reliability: the kappa statistic.
      ). κ ≤ 0 indicates no agreement, κ= 0.01–0.20 indicates poor agreement, κ=0.21–0.40 indicates fair agreement, κ= 0.41–0.60 indicates moderate agreement, κ= 0.61–0.80 indicates substantial agreement, and κ= 0.81–1.00 indicates almost perfect agreement (
      • McHugh MLJBm
      Interrater reliability: the kappa statistic.
      ). Concordance analysis between FinecareTM and sVNT and VIDAS®3 were conducted. The concordance measures included overall, positive, and negative percent agreement and Cohen's Kappa statistic. The correlation between FinecareTM and each immunoassay was examined using Pearson's correlation coefficients (r) with 95% CI. A coefficient of 0–0.19 suggests very weak correlation, 0.2–0.39 suggests weak correlation, 0.40–0.59 suggests moderate correlation, 0.6–0.79 suggests strong correlation, and 0.8–1 suggests very strong correlation (

      Swinscow TDV, Campbell MJ. Statistics at square one: Bmj London, 2002.

      ). Receiving operating characteristic (ROC) curve and Youden's index (J) were used to assess the assay threshold (cutoff indexes), identify optimized ones, and measure the area under the curve (AUC). The relation between AUC and diagnostic accuracy is direct. The larger the AUC the more accurate a test can be considered in its overall performance. Statistically, an AUC of <0.5 suggests no discrimination, 0.7–0.8 is considered acceptable, 0.8–0.9 is considered excellent, and >0.9 is considered outstanding. To determine the optimal threshold for optimal sensitivity and specificity for FinecareTM, J was calculated using the following formula: J = max (sensitivity + specificity) – 1 (
      • Fluss R
      • Faraggi D
      • BJBJJoMMiB Reiser
      Estimation of the Youden Index and its associated cutoff point.
      ,

      Unal IJC, medicine mmi. Defining an optimal cut-point value in ROC analysis: an alternative approach. Computational and mathematical methods in medicine. 2017.

      ). All statistical analyses were performed using GraphPad Prism version 9.3.0.

      3. Results

      3.1 Diagnostic Performance of FinecareTM using RT-PCR as a reference test

      The overall diagnostic performance of Finecare analyzer compared with RT-PCR is summarized in Table 1. The sensitivity and specificity of FinecareTM were 92% (95% CI: 86.44%–95.80%) and 100% (95% CI: 96.38%–100.00%), respectively. The overall percent agreement with RT-PCR was 96.4% (95% CI: 93.2%–98.1%). Cohen's Kappa statistic with RT-PCR denoted an excellent agreement with κ coefficient of 0.881 (95% CI: 0.816–0.946).
      Table 1Diagnostic assessment of Finecare for S-RBD total antibodies detection* RT-PCR was used as a reference test.
      ImmunoassayFinecareTM
      Sensitivity, % (95% CI)92% (95% CI: 86.44%–95.80%)
      Specificity, % (95% CI)100% (95% CI: 96.38%–100.00%)
      Overall agreement, % (95% CI)96.4% (95% CI: 93.2%–98.1%)
      Cohen's kappa coefficient, κ (95% CI)0.881 (95% CI: 0.816–0.946)
      ROC curves optimized cutoff index>10.70
      Sensitivity using optimized cutoff indexes, % (95% CI)94.67% (95%CI: 89.83%–97.27%)
      Specificity using optimized cutoff indexes, % (95% CI)100% (95% CI: 96.34%–100%)
      Abbreviations: ROC, receiver operating characteristic; RT-PCR, reverse transcriptase PCR; S-RBD, SARS-CoV-2 spike protein receptor-binding domain.

      3.2 Concordance assessment between FinecareTM test, sVNT, and VIDAS®3

      The tests’ concordance assessment is summarized in Table 2. The overall percent agreement for sVNT and VIDAS®3 was 92% (95% CI: 86.5–95.4) and 94.4% (95% CI: 87.6–97.6), respectively. The positive percent agreement ranged from 95.1% (88.0–98.1) for Finecare versus VIDAS®3 to 97.7% (93.5–99.2) for FinecareTM versus sVNT. The negative percent agreement ranged from 50% (95% CI: 29–71.0) for FinecareTM versus sVNT to 88.9% (95% CI: 56.5–98.0) for FinecareTM versus VIDAS®3. Cohen's Kappa statistic denoted moderate to excellent agreement and ranged between 0.557 (95% CI: 0.32–0.78) for FinecareTM/sVNT test combination and 0.731 (95% CI: 0.51–0.95) for FinecareTM /VIDAS®3 test combination. Correlation analysis of the readings obtained from FinecareTM, sVNT, and VIDAS®3 is illustrated in Figure 1. Both immunoassays showed strong correlation with FinecareTM. Pearson's correlation coefficients (r) ranged from 0.70 for FinecareTM/sVNT to 0.8 for FinecareTM/VIDAS®3.
      Table 2Concordance assessment between the FinecareTM sVNT and VIDAS®3 tests
      Compared WithOverall Percent Agreement%(95% CI)Positive Percent Agreement%(95% CI)Negative Percent Agreement%(95% CI)Cohen's Kappa Coefficientκ (95% CI)
      FinecaresVNT92.0%(86.5–95.4)97.7%(93.5–99.2)50%(29–71.0)0.558(0.32–0.78)
      VIDAS 394.4%(87.6–97.6)95.1%(88.0–98.1)88.9%(56.5–98.0)0.731(0.51–0.95)
      Abbreviations: sVNT, surrogate virus-neutralizing test.
      Figure 1
      Figure 1Correlation analysis of the assays standardized readings obtained by each immunoassay. (A) Correlation plot of FinecareTM with the sVNT. (B) Correlation plot of FinecareTM with VIDAS®3. Pearson's correlation coefficient (r) and p-value are indicated. Pearson's r of 0–0.19 is regarded as very weak, 0.2–0.39 as weak, 0.40–0.59 as moderate, 0.6–0.79 as strong, and 0.8–1 as very strong correlation, but these are rather arbitrary limits, and the context of the results should be considered. Data are presented for 150 RT-PCR–confirmed SARS-CoV-2–positive samples. BAU, binding antibody unit; RT-PCR, reverse transcriptase–PCR; sVNT, surrogate virus-neutralizing test.

      3.3 Receiver Operating Characteristic (ROC) Curve Analysis

      ROC curve analyses showed excellent performance for FinecareTM with an AUC of 0.9627 and p < 0.0001 (Figure 2). The optimized cutoff for detecting total antibodies against RBD was derived on the basis the ROC curves and the calculated J. The cutoff obtained was >10.70 compared with the manufacturer's cutoff ≥20. Applying this new cutoff showed improved sensitivity, from 92% to 94.67%, whereas the specificity remained 100% (Table 1).
      Figure 2
      Figure 2ROC curve for Finecare 2019-nCoV RBD Antibody Test. An AUC of 0.9–1.0 is considered excellent, 0.8–0.9 is considered very good, 0.7–0.8 is considered good, 0.6–0.7 is considered sufficient, 0.5–0.6 is considered bad, and <0.5 is considered not useful. AUC, area under the curve; ROC, receiver Operating Characteristic.

      4. Discussion

      This study validated the performance of Finecare™ 2019-nCoV RBD fluorescence immunoassay for detecting total antibodies against SARS-CoV2 S-RBD after infection. A panel of 150 samples collected from RT-PCR–confirmed individuals and 100 prepandemic sera were used to evaluate the assays’ performance. To our knowledge, this is the first study conducted to validate the fluorescence-LFIA-based Finecare 2019-nCoV RBD Antibody test, which marks the novelty of this research work.
      Finecare test showed high sensitivity and specificity comparable to sVNT and VIDAS®3 and other immunoassays in the market. VIDAS®3 was reported to have a sensitivity and specificity of 88.3% and 98.4%, respectively (
      • Younes S
      • Al-Jighefee H
      • Shurrab F
      • Al-Sadeq DW
      • Younes N
      • Dargham SR
      • et al.
      Diagnostic efficiency of three fully automated serology assays and their correlation with a novel surrogate virus neutralization test in symptomatic and asymptomatic SARS-COV-2 individuals.
      ). Similarly, sVNT reported a high specificity (99.9%) and sensitivity (95%–100%) (
      • Tan CW
      • Chia WN
      • Qin X
      • Liu P
      • Chen MI-C
      • Tiu C
      • et al.
      A SARS-CoV-2 surrogate virus neutralization test based on antibody-mediated blockage of ACE2–spike protein–protein interaction.
      ). It is noteworthy that our chohort samples were collected from Qatar community volunteers; thus, the cases varied between symptomatic and asymptomatic. It was noticed that of the 12 negative samples by Finecare, 8 were asymptomatic. Earlier studies reported that severity of the case affects humoral response in patients (
      • Wu J
      • Liang B
      • Chen C
      • Wang H
      • Fang Y
      • Shen S
      • et al.
      SARS-CoV-2 infection induces sustained humoral immune responses in convalescent patients following symptomatic COVID-19.
      ,
      • Zhao J
      • Yuan Q
      • Wang H
      • Liu W
      • Liao X
      • Su Y
      • et al.
      Antibody responses to SARS-CoV-2 in patients with novel coronavirus disease 2019.
      ). Previously, we reported that the sensitivity of immunoassay was higher in symptomatic patients than in the asymptomatic patient group (
      • Younes S
      • Al-Jighefee H
      • Shurrab F
      • Al-Sadeq DW
      • Younes N
      • Dargham SR
      • et al.
      Diagnostic efficiency of three fully automated serology assays and their correlation with a novel surrogate virus neutralization test in symptomatic and asymptomatic SARS-COV-2 individuals.
      ). In addition, the sensitivity of Fincare assay for estimating anti-S-RBD or neutralizing antibodies could be overestimated because the S-RBD antigen in this assay is designed to detect antibodies specific to the wild-type (Wuhan) but not to other SARS-CoV-2 variants. Some of these variants, such as Omicron, accumulate significant numbers of mutations in the S-RBD domain. These mutations might decrease the affinity of binding and neutralizing antibodies to S-RBD, which in turn affect the sensitivity of the assay.
      We evaluated the degree of correlation between FinecareTM with sVNT and VIDAS®3. A strong correlation was obtained between FinecareTM and both assays (Figure 1); however, VIDAS®3 showed a slightly higher correlation. This is because both VIDAS®3 and FinecareTM detection methods are based on enzyme-linked fluorescent and target antibodies against the S-RBD domain. This is particularly important; the FinecareTM rapid antibody test could detect neutralizing antibodies and serve as a surrogate for the advanced automated immunoassays in clinical settings to measure the humoral immune response after vaccination or infection and research context. ROC curve analysis was also performed to determine the optimal cutoff indexes for Finecare (Figure 2). The new cutoff value showed improved sensitivity without affecting the specificity. However, the cutoff values could be adjusted depending on the clinical setting or research context. For instance, in high-prevalence settings, higher thresholds may be desirable for screening purposes, whereas lower cutoff may be helpful for diagnosis purposes (
      • Ismail A
      • Shurrab FM
      • HT Al-Jighefee
      • Al-Sadeq DW
      • Qotba H
      • Al-Shaar IA
      • et al.
      Can commercial automated immunoassays be utilized to predict neutralizing antibodies after SARS-CoV-2 infection? A comparative study between three different assays. Front Biosci.
      ).
      Our study had some limitations; most of our RT-PCR samples were collected from asymptomatic individuals (Table S1), which might have underestimated the assay's sensitivity. In addition, the control group did not include samples for other coronaviruses or influenza that might cross-react with SARS-CoV-2, which could have led to an overestimated specificity.
      In conclusion, depending on characteristics of emerging variants, our data showed that Finecare 2019-RBD antibody test showed excellent performance in terms of sensitivity, specificity, and overall agreement, with RT-PCR as a reference test. In addition, in terms of correlation with the FDA-approved sVNT from GenScript and the automated analyzer VIDAS®3 from bioMérieux, FinecareTM immunoassay showed an outstanding performance in detecting total antibodies in serum samples against SARS-CoV-2. Thus, this assay could be reliable for the quantitative detection of antibodies in the vaccinated population and recovered patients.

      Conflict of interest

      GKN would like to declare that all test kits used in this study were provided as in-kind support for his laboratory to test seroprevalence of anti-SARS-CoV-2 and antibody response among vaccinated and infected individuals in Qatar.

      Funding

      This work was made possible by grant number UREP28-173-3-057 from the Qatar National Research Fund (a member of Qatar Foundation). The statements made herein are solely the responsibility of the authors.

      Acknowledgment

      We would like to thank Qatar National Research Fund (a member of Qatar Foundation) for funding this work.

      Author contributions

      Conceptualization: GKN, NL. Participant recruitment and demographic data collection: HQ. Laboratory testing: FMS, DWA, NY. Supervision: GKN. Data analysis: GKN, FMS, NY, NL. First draft writing: GKN, FMS, NL. Review and editing: LJA, GKN. Funding: GKN, LJA.

      Ethical statement

      The study was conducted according to the guidelines of the Declaration of Helsinki and was reviewed and approved by the institutional review board of Qatar University (QU-IRB 1492-E/21 and QU-IRB 1469-E/21). Informed consent was obtained from all subjects involved in the study before collecting samples.

      Data availability

      All data produced in this study are available upon reasonable request to the authors.

      Appendix. Supplementary materials

      References

        • Al-Jighefee HT
        • Yassine HM
        • Nasrallah GKJP
        Evaluation of antibody response in symptomatic and asymptomatic COVID-19 patients and diagnostic assessment of new IgM/IgG ELISA kits.
        Pathogens. 2021; 10: 161
        • Al-Thani MH
        • Farag E
        • Bertollini R
        • Al Romaihi HE
        • Abdeen S
        • Abdelkarim A
        • et al.
        Seroprevalence of SARS-CoV-2 infection in the craft and manual worker population of Qatar.
        medRxiv. 2020; https://doi.org/10.1101/2020.11.24.20237719
        • Ben-David AJESwA
        Comparison of classification accuracy using Cohen's Weighted Kappa.
        Expert Systems with Applications. 2008; 34: 825-832
        • Corman VM
        • Landt O
        • Kaiser M
        • Molenkamp R
        • Meijer A
        • Chu DK
        • et al.
        Detection of 2019 novel coronavirus (2019-nCoV) by real-time RT-PCR.
        Eurosurveillance. 2020; 252000045
        • Dortet L
        • Ronat J-B
        • Vauloup-Fellous C
        • Langendorf C
        • Mendels D-A
        • Emeraud C
        • et al.
        Evaluating 10 commercially available SARS-CoV-2 rapid serological tests by use of the STARD (Standards for Reporting of Diagnostic Accuracy Studies) method.
        Journal of clinical microbiology. 2021; 59 (-20): e02342
        • Fluss R
        • Faraggi D
        • BJBJJoMMiB Reiser
        Estimation of the Youden Index and its associated cutoff point.
        Journal of Mathematical Methods in Biosciences. 2005; 47: 458-472
        • Ismail A
        • Shurrab FM
        • HT Al-Jighefee
        • Al-Sadeq DW
        • Qotba H
        • Al-Shaar IA
        • et al.
        Can commercial automated immunoassays be utilized to predict neutralizing antibodies after SARS-CoV-2 infection? A comparative study between three different assays. Front Biosci.
        (Landmark Ed). 2021; 26 (Jul 30PMID: 34340267): 198-206https://doi.org/10.52586/4934
        • Kang K
        • Huang L
        • Ouyang C
        • Du J
        • Yang B
        • Chi Y
        • et al.
        Development, performance evaluation, and clinical application of a Rapid SARS-CoV-2 IgM and IgG Test Kit based on automated fluorescence immunoassay.
        Journal of Medical Virology. 2021; 93: 2838-2847
        • McHugh MLJBm
        Interrater reliability: the kappa statistic.
        Biochemia medica. 2012; 22: 276-282
        • Meyer B
        • Reimerink J
        • Torriani G
        • Brouwer F
        • Godeke G-J
        • Yerly S
        • et al.
        Validation and clinical evaluation of a SARS-CoV-2 surrogate virus neutralisation test (sVNT).
        Emerging microbes & infections. 2020; 9: 2394-2403
        • Nadoushan MJ
        • Ahmadi S
        • Nadoushan PJJIJoP
        Serology testing for SARS-CoV-2: Benefits and challenges.
        Iranian Journal of Pathology. 2020; 15: 154
        • Nasrallah GK
        • Dargham SR
        • Mohammed LI
        • Abu-Raddad LJJJomv
        Estimating seroprevalence of herpes simplex virus type 1 among different Middle East and North African male populations residing in Qatar.
        Journal of medical virology. 2018; 90: 184-190
        • Nasrallah GK
        • Dargham SR
        • Sahara AS
        • Elsidiq MS
        Abu-Raddad LJJJoCV. Performance of four diagnostic assays for detecting herpes simplex virus type 2 antibodies in the Middle East and North Africa.
        Journal of Clinical Virology. 2019; 111: 33-38
        • Renard N
        • Daniel S
        • Cayet N
        • Pecquet M
        • Raymond F
        • Pons S
        • et al.
        Performance Characteristics of the Vidas SARS-CoV-2 IgM and IgG Serological Assays.
        Journal of clinical microbiology. 2021; 59 (-20): e02292
        • Smatti MK
        • Nasrallah GK
        • Al Thani AA
        • Yassine HMJV
        Measuring influenza hemagglutinin (HA) stem-specific antibody-dependent cellular cytotoxicity (ADCC) in human sera using novel stabilized stem nanoparticle probes.
        Vaccine. 2020; 38: 815-821
      1. Swinscow TDV, Campbell MJ. Statistics at square one: Bmj London, 2002.

        • Syed MA
        • Al Nuaimi AS
        • Nasrallah GK
        • Althani AA
        • Yassine HM
        • Zainel AA
        • et al.
        Epidemiology of SARS-CoV2 in Qatar's primary care population aged 10 years and above.
        BMC infectious diseases. 2021; 21: 1-11
        • Tan CW
        • Chia WN
        • Qin X
        • Liu P
        • Chen MI-C
        • Tiu C
        • et al.
        A SARS-CoV-2 surrogate virus neutralization test based on antibody-mediated blockage of ACE2–spike protein–protein interaction.
        Nature biotechnology. 2020; 38: 1073-1078
        • Theel ES
        • Slev P
        • Wheeler S
        • Couturier MR
        • Wong SJ
        • KJJocm Kadkhoda
        The role of antibody testing for SARS-CoV-2: is there one?.
        Journal of clinical microbiology. 2020; 58 (-20): e00797
      2. Unal IJC, medicine mmi. Defining an optimal cut-point value in ROC analysis: an alternative approach. Computational and mathematical methods in medicine. 2017.

        • Van Walle I
        • Leitmeyer K
        • Broberg EKJE.
        Meta-analysis of the clinical performance of commercial SARS-CoV-2 nucleic acid and antibody tests up to 22 August 2020.
        MedRxiv. 2021; 262001675
        • Wu J
        • Liang B
        • Chen C
        • Wang H
        • Fang Y
        • Shen S
        • et al.
        SARS-CoV-2 infection induces sustained humoral immune responses in convalescent patients following symptomatic COVID-19.
        Nature communications. 2021; 12: 1-9
        • Yassine HM
        • Al-Jighefee H
        • Al-Sadeq DW
        • Dargham SR
        • Younes SN
        • Shurrab F
        • et al.
        Performance evaluation of five ELISA kits for detecting anti-SARS-COV-2 IgG antibodies.
        International Journal of Infectious Diseases. 2021; 102: 181-187
        • Younes S
        • Al-Jighefee H
        • Shurrab F
        • Al-Sadeq DW
        • Younes N
        • Dargham SR
        • et al.
        Diagnostic efficiency of three fully automated serology assays and their correlation with a novel surrogate virus neutralization test in symptomatic and asymptomatic SARS-COV-2 individuals.
        Microorganisms. 2021; 9: 245
        • Zhao J
        • Yuan Q
        • Wang H
        • Liu W
        • Liao X
        • Su Y
        • et al.
        Antibody responses to SARS-CoV-2 in patients with novel coronavirus disease 2019.
        Clinical infectious diseases. 2020; 71: 2027-2034
        • Zhu F
        • Althaus T
        • Tan CW
        • Costantini A
        • Chia WN
        • Chau NVV
        • et al.
        WHO international standard for SARS-CoV-2 antibodies to determine markers of protection.
        The Lancet Microbe. 2021;